CVE Catalog

CVE-2014-5406

High
Published: Translated: NVD NIST

Summary

The Hospira LifeCare PCA Infusion System before version 7.0 does not validate network traffic associated with sending a drug library, software update, or configuration change. This allows remote attackers to modify settings or medication data via packets on TELNET, HTTP, HTTPS, or UPNP ports.

Risk Assessment

Attackers can remotely change critical settings of the infusion system, potentially leading to incorrect medication dosages, posing serious risks to patients.

Recommendation

It is recommended to update the system to version 7.0 or later and implement appropriate network security measures to restrict access to TELNET, HTTP, HTTPS, and UPNP ports.

Other vulnerabilities in Hospira LifeCare PCA Infusion System

See all
Original NVD description (English source)

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Vulnerability data from NVD (NIST) · CISA KEV · EPSS